Home    中文  
 
  • Search
  • lucene Search
  • Citation
  • Fig/Tab
  • Adv Search
Just Accepted  |  Current Issue  |  Archive  |  Featured Articles  |  Most Read  |  Most Download  |  Most Cited

Chinese Journal of Clinical Laboratory Management(Electronic Edition) ›› 2020, Vol. 08 ›› Issue (01): 31-35. doi: 10.3877/cma.j.issn.2095-5820.2020.01.006

Special Issue:

• Quality Control • Previous Articles     Next Articles

Establishment and discussion of inter-room comparison system of clinical laboratory quality control data based on capability index

Ran Tao1, Wenjing Yuan2, Jianbo Chen1, Nanxun Mo1, Jun He1, Yating Chen1, jing Feng3, Chaohui Hu1, Xiaoyan Deng2,()   

  1. 1. Guangzhou Kingmed Center for Clinical Laboratory, Laboratory diagnosis Department, Guangzhou 510005, China; Guangzhou Medical University , Guangzhou 510006, China
    2. Guangzhou Medical University , Guangzhou 510006, China
    3. Southern Medical University Affiliated Fengxian Hospital, Laboratory Medicine, Shanghai 201400, China
  • Received:2019-08-21 Online:2020-02-28 Published:2020-02-28
  • Contact: Xiaoyan Deng
  • About author:
    Corresponding author: Deng Xiaoyan, Email:

Abstract:

Objective

To establish laboratory quality control comparison system based on capability index.

Method

The quality control system is configured with comparison groups and selected rules. The capacity index is calculated after the quality control data is entered. The minimum capacity index of the same project of each laboratory is transmitted to the comparison system every ten days, and then compared with the absolute standard, the relative standard and the historical results respectively. The comparison results of each laboratory are gained to help the laboratory improve the quality control.

Results

The quality control comparison system is established. According to the WS/T403-2012 standard, 97% of the compared batches show a good or or above precision. However, five batches of total cholesterol and glucose show a general precision, which needs to be improved individually. Moreover, 95.8% of the compared batches show good or above accuracy and the system error is the main factor leading to the degradation of enzyme project performance.

Conclusion

The comparison system based on capability index can helps managers to identify potential quality control hidden dangers in terms of precision and accuracy, and provide performance improvement suggestions for various testing items according to the current technical level, and ensure the safety of patients’ results.

Key words: Clinical laboratory techniques, Quality improvement, Clinical laboratory information systems