Construction strategy and optimization of automatic review system in clinical chemistry laboratory

Wenhai Jiang1, Jiong Wu2, Yiqing Yin1,(), Baishen Pan2   

  1. 1.Computer Network Center, Zhongshan Hospital Affiliated to Fudan University, Shanghai 200032, China
    2.Department of Laboratory,Zhongshan Hospital Affiliated to Fudan University, Shanghai 200032, China
  • Received:2015-05-20 Published:2015-05-28
  • Corresponding author: Yiqing Yin
引用本文:

蒋文海, 吴炯, 阴忆青, 潘柏申. 临床化学实验室自动审核系统搭建策略及优化[J/OL]. 中华临床实验室管理电子杂志, 2015, 03(02): 97-101.

Wenhai Jiang, Jiong Wu, Yiqing Yin, Baishen Pan. Construction strategy and optimization of automatic review system in clinical chemistry laboratory[J/OL]. Chinese Journal of Clinical Laboratory Management(Electronic Edition), 2015, 03(02): 97-101.

目的

建立能满足临床数据审核需求的安全有效的临床化学实验室自动审核系统。

方法

复旦大学附属中山医院检验科以罗氏中间体软件IT3000为核心,整合本院自行开发与维护的实验室信息系统(laboratory information system, LIS),并运用分析仪器搭建临床化学实验室自动审核系统,以评价2012年至2015年期间自动化审核系统使用前后周转时间(turn around time,TAT)合格率变化趋势;同时,分析自动审核系统使用前后降低检测报告的差错率(以报告解除确认率表示)的效果。

结果

在2012年1至3月搭建起由LIS、中间体软件、检测仪器和数据审核人员四大部分组成的本院自动审核系统;同时,建立项目匹配、仪器报警、多次结果比较、限值管理、质控规则、逻辑规则六层数据审核策略及审核报警警示系统。用于分析自动审核系统使用前2012年第一季度TAT超时率为0.281% (238/84729),2013年、2014年、2015年第一季度超时率分别为0.058% (73/126605)、0.051%(86/169403)、0.056% (96/171821),较2012年第一季度分别下降42%、82%、80%。未使用自动化审核系统的2012年第一季度报告解除确认率为0.16% (135/84729),2013年、2014年、2015年第一季度解除确认率分别为0.07% (87/126605)、0.03% (51/169403)、0.08% (142/171821),较2012年第一季度分别下降57%、81%、48%。

结论

临床化学实验室自动审核系统有效的增强工作效率,提高了TAT合格率,降低了报告差错率。

Objective

To establish a safe and effective Auto-validation system for meeting the requirements of clinical data validation in clinical chemistry laboratories.

Methods

The Clinical Laboratory of Zhongshan Hospital, affiliated to Fudan University used Roche′s middleware IT3000 and integrated a series of other software and hardware [such as Laboratory information system(LIS), analyzers, Data flow, quality control and sample flow monitoring] to create an auto-validation system. The trend in turnaround time(TAT) pass rate from 2012 to 2015, before and after the adoption of the auto-validation system, and the reduction of error rates were evaluated.

Results

An auto-validation system had been adopted since Jan. 2012, which comprises of LIS(developed and maintained by Zhongshan Hospital in 3 months), middleware, clinical analyzers and data validation staff members. The auto-validation and alert strategies includes six elements: test matching, instrument alarm,multiple results comparison, limit value management, quality control rules, and logical rules. Before using the autovalidation system, the TAT overtime rate in the first quarter of 2012 was 0.281% (238/84729), while that in the same quarter of 2013, 2014, and 2015 was 0.058% (73/126605), 0.051% (86/169403), and 0.056% (96/171821)respectively, which represented a decrease of 42%, 82%, and 80%. The confirmation rate of report cancellation in the first quarter of 2012 was 0.16% (135/84729), and dropped 57%, 81%, and 48% in 2013, 2014, and 2015 to 0.07% (87/126605),0.03% (51/169403),0.08% (142/171821).

Conclusion

The auto-validation system has highly sped up the result validation process and effectively reduced report error rate.

图1 我院自动化审核系统模块功能设置
图2 复旦大学附属中山医院检验科自动化审核系统对数据审核的五层策略
表1 自动生化分析仪中部分项目血清指数报警限值
表2 部分常用检验项目多次结果比较差值校验范围及限值审核范围
表3 中山医院检验科自动审核逻辑规则
图3 本院自建自动化审核系统2012年至2015年的TAT超时率和报告解除确认率图
1
Gomez-Ruben R, Alvarez V, Ventura M, et al. Current status of verification practics in clinical biochemistry in Spain [J]. Clin Chem Lab Med, 2013,51(9):1739-1746.
2
CLSI. CLSI AUTO10-A(ISBN 1-56238-620-4) Auto verification of clinical laboratory test results; approved guideline. CLSI, 2006.
3
Torke N, Boral L, Nguyen T, et al. Process improvement and operational efficiency through test result autoverification [J]. Clin Chem, 2005,51(12):2406-2408.
4
Guidi GC, Poli G, Bassi A, et al. Development and implementation of an automatic system for verification, validation and delivery of laboratory test results [J]. Clin Chem Lab Med, 2009,97(11):1355-1360.
5
Corberand JX, Rogari E, Laharrague P, et al. Computer-assisted validation system applied to hematology : Valab-haemato[J]. Ann Biol Clin, 1994,52(6):447-450.
6
Oosterhuis WP, Ulenkate HJ, Goldschmidt HM. Evaluation of LabRespond, a new automated validation system for clinical laboratory[J]. Clin Chem, 2000,46(11):1811-1817.
7
姜傥. 全面实验室自动化与自动化管理理念[J]. 中华医学检验杂志,2004,27(3):129-131.
8
陈文, 袁劲松, 钱瑛, 等. 在LIS支持下实验室自动化系统流程的规范和优化[J]. 临床检验杂志, 2011,29(7):495.
9
周智刚, 张建臣, 郭长友, 等. 计算机实验室软件的自动化管理[J].实验室研究与探索, 2006,25(11):1376-1379.
[1] 张占卿, 陆伟, 周新兰, 王雁冰, 丁荣蓉, 黄丹, 李秀芬. Lumipulse HBsAg-HQ与Architect HBsAg-QT定量的一致性及其预测慢性乙型肝炎肝组织病理状态的效能[J/OL]. 中华实验和临床感染病杂志(电子版), 2017, 11(05): 447-454.
[2] 倪兆慧, 金海姣. 自动化腹膜透析的新应用[J/OL]. 中华肾病研究电子杂志, 2015, 04(01): 10-13.
[3] 赵平, 廖昌友, 丁伟. 射频识别技术在临床病理科全流程管理中的应用[J/OL]. 中华临床实验室管理电子杂志, 2023, 11(04): 193-199.
[4] 甄拴平, 王华, 朱海鹏, 豆淑媛, 赵雪莲. 基层医院检验危急值管理的实践与探讨[J/OL]. 中华临床实验室管理电子杂志, 2023, 11(01): 1-4.
[5] 胡朝友, 张星晨, 吴海珍, 常章梅. 基于Java和MySQL软件的实验室菌种管理系统设计与应用[J/OL]. 中华临床实验室管理电子杂志, 2020, 08(01): 41-47.
[6] 陶然, 袁文静, 陈建波, 莫南勋, 何君, 程雅婷, 冯景, 胡朝晖, 邓小燕. 基于能力指数的临床实验室质量控制数据室间比对系统的建立与探讨[J/OL]. 中华临床实验室管理电子杂志, 2020, 08(01): 31-35.
[7] 张磊, 朱庆义. 感染性疾病诊断新技术的应用[J/OL]. 中华临床实验室管理电子杂志, 2017, 05(02): 89-92.
[8] 戴涵清, 俞磊, 李艺, 陆胜, 职庭帆, 杨光, 周学迅, 王雪琦. 自动化存储管理系统在生物样本库领域的应用及展望[J/OL]. 中华临床实验室管理电子杂志, 2017, 05(01): 46-50.
[9] 裴世静, 刘金萍, 马斌国. 山西省医疗机构临床实验室基本情况调查分析[J/OL]. 中华临床实验室管理电子杂志, 2016, 04(01): 45-48.
[10] 邱玲. ISO 15189评审自动生化分析仪使用中的常见不符合项及解决方案[J/OL]. 中华临床实验室管理电子杂志, 2015, 03(04): 198-201.
[11] 杨汝, 魏军, 李锋, 王利新, 苏明, 王青. 办公自动化系统在临床实验室的应用[J/OL]. 中华临床实验室管理电子杂志, 2015, 03(01): 13-16.
[12] 康凤凤, 郦卫星. 自动化免疫血液分析仪的使用前确认[J/OL]. 中华临床实验室管理电子杂志, 2014, 02(03): 175-178.
[13] 陈佑平, 张真路, 李梦雅. 心脑血管病自动化风险评估系统的建立及其初步应用效果分析[J/OL]. 中华临床实验室管理电子杂志, 2014, 02(02): 120-127.
[14] 杨启文, 徐英春. 临床微生物实验室自动化现状与进展[J/OL]. 中华临床实验室管理电子杂志, 2014, 02(01): 4-10.
[15] 范欣, 邱玲, 徐英春. 实验室全面自动化:细菌与真菌实验室面临的机遇与挑战[J/OL]. 中华临床实验室管理电子杂志, 2014, 02(01): 1-3.
阅读次数
全文


摘要